Customization: | Available |
---|---|
After-sales Service: | Lifelong |
Warranty: | 1 Year |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
It is used to test the burning performance of the medical mask after it contacts the flame at a certain linear speed.
Standard
GB 19083-2010,YY 0469-2011,GB 2626-2006,etc
Feature
1,Adopt all stainless steel material appearance to prevent smoking.
2,Industrial color touch screen display control, durable, easy to operate and simple.
3,The fixture of the mask is a metal human head mold, which can fully simulate the actual use of the mask.
4,The burner is equipped with a combustible gas interface such as propane (liquefied gas)
5,The height of the die can be adjusted.
6,Automatic timing positioning of burner.
7,Combustion time, flame retardant time are automatically recorded, digital display.
8,Equipped with flame temperature measuring probe, real-time temperature display.
9,The linear velocity can be set:30~99mm/s.
Parameter
1,Measurement Range:
afterflame time:0.1~999.9s accuracy0.1s
afterglow time:0.1~999.9s acccuracy0.1s
2,Preheating time:1~60s
3,The flame height can be adjusted(Factory adjustment at 40±4mm)
4,The diameter of the flame temperature measuring probe is 1.5mm
5,The flame temperature is (800±50) ºC at a height of (20±2) mm from the top of the burner.
6,The distance between the top of the burner and the lowest point of the mask is (20±2) mm.
7,Metal die wall thickness of 2 ~ 3mm, flame resistance and combustion resistance.
8,Afterburn time and smolder time timer:0~99.9S,accuracy:±0.1s
9,Use gas: propane gas in a tank.
10,weight:about 30kg
11,power:AC220V 50Hz 50W
11,power:AC220V 50Hz 50W |